Product Description
-
Diagnostic X-Ray High-Voltage Generator for Stationary DR System
The DMP-100 is a medical X-ray generator designed for imaging applications. With 450-kHz high-frequency performance and 3-phase adaptive input voltage, the DMP-100 improves image quality by effectively reducing ripple and noise. To satisfy different X-ray system integration needs, this unit also offers 2-speed motor-starter support, 1 phase input power and supplies power to AEC/DAP peripherals for optimized configuration. It applies a digital control circuit to enhance performance and accuracy, making it a powerful core for stationary DR/CR systems.
Product Spec
-
-
32kW
- Input Voltage 230VAC ± 10%, 1Ф 230/400 – 277/480VAC ±10%, 3Ф
- Input Current 260A ± 10% @ 230VAC, 80A ± 10% @ 400VAC, 70A ± 10% @ 480VAC
- Input Frequency 50Hz / 60Hz
- kV Range 40kV - 150kV, 1kV / Step
- kV Accuracy ± (5% of preset value + 1kV)
- Rise Time < 1.5ms (10% - 90%)
- mA Range 10mA - 400mA
- mA Accuracy ± (5% of preset value + 1mA)
- ms Range 1ms - 10000ms
- ms Accuracy ± (10% of preset value + 1ms)
- mAs Range 0.1mAs - 500mAs
- mAs Accuracy ± (5% of preset value + 0.2mAs)
- Max. Output Power 320mA /100kV
- External Output Voltage 24VDC ± 8% / 2A, 24VAC ± 10% / 6A, 110VAC ± 10% / 0.8A or 220VAC ± 10% / 0.8A
- Motor Starter Dual Speed Auto Control (High Speed Operation only available at 400VAC - 480VAC, 3Ф models)
- Support Peripherals AEC / DAP / DR
- Dimension 654 mm* 352 mm* 482 mm (L*W*H)
- Weight 75 kg
- Console Optional, APR: 64 Preset for User Programmable
- Safety IEC 60601-1 2nd edition, IEC 60601-1 3.1rd edition and IEC 60601-1 3.2rd edition, IEC 60601-2-54
- EMC IEC 60601-1-2 4ᵗʰ edition compliance
- Certificate CE MDD(0197) / CE EMC / CB
-
40kW
- Input Voltage 230VAC ± 10%, 1Ф 230/400 – 277/480VAC ±10%, 3Ф
- Input Current 325A± 10% @ 230VAC, 100A ± 10% @ 400VAC, 88A ± 10% @ 480VAC
- Input Frequency 50Hz / 60Hz
- kV Range 40kV - 150kV, 1kV / Step
- kV Accuracy ± (5% of preset value + 1kV)
- Rise Time < 1.5ms (10% - 90%)
- mA Range 10mA - 500mA
- mA Accuracy ± (5% of preset value + 1mA)
- ms Range 1ms - 10000ms
- ms Accuracy ± (10% of preset value + 1ms)
- mAs Range 0.1mAs - 630mAs
- mAs Accuracy ± (5% of preset value + 0.2mAs)
- Max. Output Power 400mA /100kV
- External Output Voltage 24VDC ± 8% / 2A, 24VAC ± 10% / 6A, 110VAC ± 10% / 0.8A or 220VAC ± 10% / 0.8A
- Motor Starter Dual Speed Auto Control (High Speed Operation only available at 400VAC - 480VAC, 3Ф models)
- Support Peripherals AEC / DAP / DR
- Dimension 654 mm* 352 mm* 482 mm (L*W*H)
- Weight 75 kg
- Console Optional, APR: 64 Preset for User Programmable
- Safety IEC 60601-1 2nd edition, IEC 60601-1 3.1rd edition and IEC 60601-1 3.2rd edition, IEC 60601-2-54
- EMC IEC 60601-1-2 4ᵗʰ edition compliance
- Certificate CE MDD(0197) / CE EMC / CB
-
50kW
- Input Voltage 230/400 – 277/480VAC ±10%, 3Ф
- Input Current 115A ± 10% @400VAC, 100A ± 10% @ 480VAC
- Input Frequency 50Hz / 60Hz
- kV Range 40kV - 150kV, 1kV / Step
- kV Accuracy ± (5% of preset value + 1kV)
- Rise Time < 1.5ms (10% - 90%)
- mA Range 10mA - 630mA
- mA Accuracy ± (5% of preset value + 1mA)
- ms Range 1ms - 10000ms
- ms Accuracy ± (10% of preset value + 1ms)
- mAs Range 0.1mAs - 1250mAs
- mAs Accuracy ± (5% of preset value + 0.2mAs)
- Max. Output Power 500mA /100kV
- External Output Voltage 24VDC ± 8% / 2A, 24VAC ± 10% / 6A, 110VAC ± 10% / 0.8A or 220VAC ± 10% / 0.8A
- Motor Starter Dual Speed Auto Control (High Speed Operation only available at 400VAC - 480VAC, 3Ф models)
- Support Peripherals AEC / DAP / DR
- Dimension 654 mm* 352 mm* 482 mm (L*W*H)
- Weight 75 kg
- Console Optional, APR: 64 Preset for User Programmable
- Safety IEC 60601-1 2nd edition, IEC 60601-1 3.1rd edition and IEC 60601-1 3.2rd edition, IEC 60601-2-54
- EMC IEC 60601-1-2 4ᵗʰ edition compliance
- Certificate CE MDD(0197) / CE EMC / CB
-
56kW
- Input Voltage 230/400 – 277/480VAC ±10%, 3Ф
- Input Current 129A ± 10% @400VAC, 112A ± 10% @480VAC
- Input Frequency 50Hz / 60Hz
- kV Range 40kV - 150kV, 1kV / Step
- kV Accuracy ± (5% of preset value + 1kV)
- Rise Time < 1.5ms (10% - 90%)
- mA Range 10mA - 800mA
- mA Accuracy ± (5% of preset value + 1mA)
- ms Range 1ms - 10000ms
- ms Accuracy ± (10% of preset value + 1ms)
- mAs Range 0.1mAs - 1250mAs
- mAs Accuracy ± (5% of preset value + 0.2mAs)
- Max. Output Power 560mA /100kV
- External Output Voltage 24VDC ± 8% / 2A, 24VAC ± 10% / 6A, 110VAC ± 10% / 0.8A or 220VAC ± 10% / 0.8A
- Motor Starter Dual Speed Auto Control (High Speed Operation only available at 400VAC - 480VAC, 3Ф models)
- Support Peripherals AEC / DAP / DR
- Dimension 654 mm* 352 mm* 482 mm (L*W*H)
- Weight 75 kg
- Console Optional, APR: 64 Preset for User Programmable
- Safety IEC 60601-1 2nd edition, IEC 60601-1 3.1rd edition and IEC 60601-1 3.2rd edition, IEC 60601-2-54
- EMC IEC 60601-1-2 4ᵗʰ edition compliance
- Certificate CE MDD(0197) / CE EMC / CB
-
65kW
- Input Voltage 230/400 – 277/480VAC ±10%, 3Ф
- Input Current 150A ± 10% @400VAC, 120A ± 10% @480VAC
- Input Frequency 50Hz / 60Hz
- kV Range 40kV - 150kV, 1kV / Step
- kV Accuracy ± (5% of preset value + 1kV)
- Rise Time < 1.5ms (10% - 90%)
- mA Range 10mA - 800mA
- mA Accuracy ± (5% of preset value + 1mA)
- ms Range 1ms - 10000ms
- ms Accuracy ± (10% of preset value + 1ms)
- mAs Range 0.1mAs - 1250mAs
- mAs Accuracy ± (5% of preset value + 0.2mAs)
- Max. Output Power 800mA /80kV
- External Output Voltage 24VDC ± 8% / 2A, 24VAC ± 10% / 6A, 110VAC ± 10% / 0.8A or 220VAC ± 10% / 0.8A
- Motor Starter Dual Speed Auto Control (High Speed Operation only available at 400VAC - 480VAC, 3Ф models)
- Support Peripherals AEC / DAP / DR
- Dimension 654 mm* 352 mm* 482 mm (L*W*H)
- Weight 75 kg
- Console Optional, APR: 64 Preset for User Programmable
- Safety IEC 60601-1 2nd edition, IEC 60601-1 3.1rd edition and IEC 60601-1 3.2rd edition, IEC 60601-2-54
- EMC IEC 60601-1-2 4ᵗʰ edition compliance
- Certificate CE MDD(0197) / CE EMC / CB
-
71kW
- Input Voltage 230/400 – 277/480VAC ±10%, 3Ф
- Input Current 164A ± 10% @400VAC, 132A ± 10% @480VAC
- Input Frequency 50Hz / 60Hz
- kV Range 40kV - 150kV, 1kV / Step
- kV Accuracy ± (5% of preset value + 1kV)
- Rise Time < 1.5ms (10% - 90%)
- mA Range 10mA - 800mA
- mA Accuracy ± (5% of preset value + 1mA)
- ms Range 1ms - 10000ms
- ms Accuracy ± (10% of preset value + 1ms)
- mAs Range 0.1mAs - 1250mAs
- mAs Accuracy ± (5% of preset value + 0.2mAs)
- Max. Output Power 710mA /100kV
- External Output Voltage 24VDC ± 8% / 2A, 24VAC ± 10% / 6A, 110VAC ± 10% / 0.8A or 220VAC ± 10% / 0.8A
- Motor Starter Dual Speed Auto Control (High Speed Operation only available at 400VAC - 480VAC, 3Ф models)
- Support Peripherals AEC / DAP / DR
- Dimension 654 mm* 352 mm* 482 mm (L*W*H)
- Weight 75 kg
- Console Optional, APR: 64 Preset for User Programmable
- Safety IEC 60601-1 2nd edition, IEC 60601-1 3.1rd edition and IEC 60601-1 3.2rd edition, IEC 60601-2-54
- EMC IEC 60601-1-2 4ᵗʰ edition compliance
- Certificate CE MDD(0197) / CE EMC / CB
-
80kW
- Input Voltage 230/400 – 277/480VAC ±10%, 3Ф
- Input Current 185A ± 10% @400VAC, 150A ± 10% @480VAC
- Input Frequency 50Hz / 60Hz
- kV Range 40kV - 150kV, 1kV / Step
- kV Accuracy ± (5% of preset value + 1kV)
- Rise Time < 1.5ms (10% - 90%)
- mA Range 10mA - 1000mA
- mA Accuracy ± (5% of preset value + 1mA)
- ms Range 1ms - 10000ms
- ms Accuracy ± (10% of preset value + 1ms)
- mAs Range 0.1mAs - 1250mAs
- mAs Accuracy ± (5% of preset value + 0.2mAs)
- Max. Output Power 800mA /100kV
- External Output Voltage 24VDC ± 8% / 2A, 24VAC ± 10% / 6A, 110VAC ± 10% / 0.8A or 220VAC ± 10% / 0.8A
- Motor Starter Dual Speed Auto Control (High Speed Operation only available at 400VAC - 480VAC, 3Ф models)
- Support Peripherals AEC / DAP / DR
- Dimension 654 mm* 352 mm* 482 mm (L*W*H)
- Weight 75 kg
- Console Optional, APR: 64 Preset for User Programmable
- Safety IEC 60601-1 2nd edition, IEC 60601-1 3.1rd edition and IEC 60601-1 3.2rd edition, IEC 60601-2-54
- EMC IEC 60601-1-2 4ᵗʰ edition compliance
- Certificate CE MDD(0197) / CE EMC / CB
-
82kW
- Input Voltage 230/400 – 277/480VAC ±10%, 3Ф
- Input Current 190A ± 10% @400VAC, 154A ± 10% @480VAC
- Input Frequency 50Hz / 60Hz
- kV Range 40kV - 150kV, 1kV / Step
- kV Accuracy ± (5% of preset value + 1kV)
- Rise Time < 1.5ms (10% - 90%)
- mA Range 10mA - 1000mA
- mA Accuracy ± (5% of preset value + 1mA)
- ms Range 1ms - 10000ms
- ms Accuracy ± (10% of preset value + 1ms)
- mAs Range 0.1mAs - 1250mAs
- mAs Accuracy ± (5% of preset value + 0.2mAs)
- Max. Output Power 800mA /103kV
- External Output Voltage 24VDC ± 8% / 2A, 24VAC ± 10% / 6A, 110VAC ± 10% / 0.8A or 220VAC ± 10% / 0.8A
- Motor Starter Dual Speed Auto Control (High Speed Operation only available at 400VAC - 480VAC, 3Ф models)
- Support Peripherals AEC / DAP / DR
- Dimension 654 mm* 352 mm* 482 mm (L*W*H)
- Weight 75 kg
- Console Optional, APR: 64 Preset for User Programmable
- Safety IEC 60601-1 2nd edition, IEC 60601-1 3.1rd edition and IEC 60601-1 3.2rd edition, IEC 60601-2-54
- EMC IEC 60601-1-2 4ᵗʰ edition compliance
- Certificate CE MDD(0197) / CE EMC / CB
-
-
Document Download