Product Description
-
Low-bed X-ray generator for diagnosis involving mobility-impaired patients
The DMP-200 applies a flat design for medical low-bed DR systems. The unit is useful for applications where space constraints are a major consideration, such as when performing diagnosis on elderly people and those with disabilities. With 450-kHz high-frequency performance, it improves image quality by effectively reducing ripple and noise. Combined with a universal 100–240 VAC input standard, 2-speed motor starter, and AEC/DAP peripheral connectivity, it satisfies the needs of medical equipment system manufacturers.
Product Spec
-
-
32kW
- Input Voltage 100 VAC - 240 VAC 1Ф
- Input Current Range 15 A - 10 A
- Input Frequency 50 Hz / 60 Hz
- kV Range 40 kV – 150 kV, 1 kV / Step
- kV Accuracy ± (5% of preset value + 1 kV)
- Rise Time < 1.5 ms (10% - 90%)
- mA Range 10 mA - 500 mA, R’20
- mA Accuracy +/- (5% of preset value + 1 mA)
- ms Range 1 ms – 6300 ms, R’20
- ms Accuracy ± (10% of preset value + 1 ms)
- mAs 42 kV / 100 mAs, 125 kV / 32 mAs, 32 kW / 100 ms max, R’20
- mAs Accuracy ± (5% of preset value + 0.2 mAs)
- Max. Output Power 320 mA / 100 kV
- External Output Voltage 24 VDC / 2 A, 24 AC / 6 A, 110 VAC / 2 A or 220
- Motor Starter Low Speed
- Support Peripherals AEC / DAP, Optional
- Dimension (W x H x D) 630 mm x 400 mm x 370 mm
- Weight 60 kg
- Console Optional, APR: 192 Preset for User Programmable
- Safety IEC 60601-1 2nd edition, IEC 60601-1 3.1rd edition and IEC 60601-1 3.2rd edition, IEC 60601-2-54
- EMC IEC 60601-1-2 4ᵗʰ edition compliance
- Certificate CB
-
40kW
- Input Voltage 100 VAC - 240 VAC 1Ф
- Input Current Range 15 A - 10 A
- Input Frequency 50 Hz / 60 Hz
- kV Range 40 kV – 150 kV, 1 kV / Step
- kV Accuracy ± (5% of preset value + 1 kV)
- Rise Time < 1.5 ms (10% - 90%)
- mA Range 10 mA - 500 mA, R’20
- mA Accuracy +/- (5% of preset value + 1 mA)
- ms Range 1 ms – 6300 ms, R’20
- ms Accuracy ± (10% of preset value + 1 ms)
- mAs 140 mAs (40 kW), 280 mAs max, R’20
- mAs Accuracy ± (5% of preset value + 0.2 mAs)
- Initial Charging Time 35 sec
- Recharge Time after Max. mAs Output 18 sec
- Max. Output Power 400 mA / 100 kV
- External Output Voltage 24 VDC / 2 A, 24 AC / 6 A, 110 VAC / 2 A or 220
- Motor Starter Low Speed
- Support Peripherals AEC / DAP, Optional
- Capacitance 200,000 mF
- Dimension (W x H x D) Generator: 630 mm x 370 mm x 400 mm, Capacitor Bank : 503 mm x 359 mm x 195 mm
- Weight Generator: 60 kg, Capacitor Bank: 25.5 kg
- Console Optional, APR: 192 Preset for User Programmable
- Safety IEC 60601-1 2nd edition, IEC 60601-1 3.1rd edition and IEC 60601-1 3.2rd edition, IEC 60601-2-54
- EMC IEC 60601-1-2 4ᵗʰ edition compliance
- Certificate CB
-
-